Show all Refrigerators

Ultra-low temperature freezer -86℃ (188L)

Safety:

  • Fault alarm including high and low temperature, power failure, sensor error, filter and condenser cleaning, abnormal voltage, thermostat failure, low battery
  • Able to issue two types of alarms: audible buzzer and visible flashing light
  • Remote alarm

Scope of use

Suitable for low-temperature testing of special materials and preservation of plasma, biological materials, vaccines, reagents, biological products, chemical reagents, strains, biological samples, etc.

Energy Saving and Energy Efficiency:

  • High-performance VIP insulation panels (Vacuum Insulation Panels) minimize heat transfer from the cabinet and improve temperature stability
  • Unique feature of internal door insulation design with four separate storage compartments, minimizes frost formation inside the chamber
  • Unique door seal design for minimal cold temperature loss when opening the door
  • Refrigerant with better efficiency and lower energy consumption, safe for the environment

Key Features:

  • Microprocessor system designed with 1℃ increments
  • Configurable high and low temperature alarms
  • Automatic filter cleaning alarm and sensor error notification
  • Adjustable height of storage shelves
Technical specifications
Model ULF86-190Е
Usable volume (l)188
Internal dimensions (W x D x H)465 х 465 х 830 mm
External dimensions (W x D x H)830 x 810 x 1607.5 mm
DisplayDigital
Net/gross weight (kg)207/220
Package dimensions (W x D x H)890 х 920 х 1820 mm
Productivity
Operating range -40…-86 ℃
SensorРТС
Controller typeMicroprocessor
Structure
Inner materialStainless steel, grade 304
Outer materialGalvanized steel with antimicrobial powder coating
Interior doorsStainless steel, grade 304
Shelves/Drawers (qty)Shelf/2
WheelsStandard equipment
Cooling system
RefrigerantR4001 (CFC-free)
DefrostingManual
Compressor/qtyCUBIGEL/1
Electricity
Noise level (dB) 52
Voltage/Frequency220V ± 10%/50 Hz
Power (W)4660
Security system
USB-portStandard equipment
Remote alarm portStandard equipment
Test port (qty)/Diameter (mm)1/25
AlarmsHigh and low temperature, high and low ambient temperature, voltage failure, low battery, power failure, UPS for alarm (72 hours), door open

    Get in touch with us